Session aimTest MARTe 1 chopper acquisition modules and perform PF magnetic field tests
Session day2024-12-20Operatorfernando
Session IDISTTOK_20241220_1Coordinatorrhenriques

Shot Time Operator Iplasma(kA) n(1018m-3) Duration (ms) Pre/post shot comments
516302024-12-20 17:39:40.289fernando000Primary field only with triangular waveform | OK, but strange primary current behavior (maybe related with MARTe operation modes?). Good for purpose anyway
516292024-12-20 17:33:15.90fernando000Primary field only with triangular waveform | forgot to switch ON 2nd level of breaker protection
516282024-12-20 17:28:58.003fernando000Primary field only with triangular waveform | forgot to switch ON capacitor bank charger
516272024-12-20 17:19:20.10fernando000Horizontal field only with triangular waveform | Good, better field penetration than in vertical
516262024-12-20 17:15:14.768fernando000Horizontal field only with triangular waveform | PS breaker was open
516252024-12-20 16:55:32.754fernando000Vertical Field only with triangular waveforms, MARTe 2 connections reconnected | Very interesting results
516242024-12-20 16:39:33.945fernando000Repeat with connection to MARTe2 removed aiming at EO reduction on MARTe acq. modules | Did not change much, exept in one or two
516232024-12-20 16:14:44.121fernando000Vertical field only (200 A) 13 ms long | OK, only seen on coil no. 2?, 3, 4, 5, 6, 8, and 9?
516222024-12-20 15:37:10.588rhenriques000Test all chopper acquisition modules, square wave, f=33.3(3)Hz, +/-1V | Good channels: 129 to 136, 138; Bad: 137, 139 and 140