Session aimPulses for acquistions tests and calibration
Session day2025-03-17Operatorrhenriques
Session IDISTTOK_20250317_1Coordinatorrhenriques

Shot Time Operator Iplasma(kA) n(1018m-3) Duration (ms) Pre/post shot comments
519702025-03-17 18:22:43.394rhenriques000Repeat with 0 V (shorted) | OK
519692025-03-17 18:18:04.849rhenriques000Repeat with +4.096 V | OK
519682025-03-17 18:12:00.692rhenriques000Calibration of B3 all chs. (1-32) with -4.096 V | OK
519672025-03-17 18:01:48.766rhenriques000Repeat to test amp no. 4 and 7 with different acquistion cable | OK
519662025-03-17 17:57:17.907rhenriques000Repeat testing HIBD amps no. 5 and 6 | OK
519652025-03-17 17:51:24.439rhenriques000Repeat with swapped channels 3 and 4 as 4 was did not work in previous pulse | OK, no. 4 did not work, may be the cable
519642025-03-17 17:46:18.888rhenriques000Repeat, now with amplifiers 3 and 4 | OK
519632025-03-17 17:40:43.906rhenriques000Repeat | OK
519622025-03-17 17:31:49.518rhenriques000Repeat | OK
519612025-03-17 17:27:28.474rhenriques000Repeat just for HIBD acquisition tests | OK
519602025-03-17 17:06:19.834rhenriques000Test all RT1 channels with 13 Vpp sine wave, HIBD acquisition tests continues | OK, but signals generated a HARD stop in MARTe
519592025-03-17 16:57:42.422rhenriques000Calibration of RT2 chs. 17-28 with 0 V (shorted), HIBD tests on B3 ch. 9-16 and 25-32 | OK
519582025-03-17 16:42:09.673rhenriques000Calibration of RT2 chs. 17-28 with -4.096 V, HIBD tests on B3 ch. 9-16 and 25-32 | OK
519572025-03-17 16:24:59.993rhenriques000Calibration of RT2 chs. 17-28 with +4.096 V, HIBD tests on B3 ch. 17-32 | OK
519562025-03-17 15:51:08.263rhenriques000Calibration of RT2 chs. 17-24 with -4.096 V, RT2 chs. 9 and 16 shorted (0 V), HIBD tests on B3 ch. 17-32 | OK
519552025-03-17 15:12:25.689rhenriques000Repeat with repaired ADC module no. 28 and inspected module 12 (which was working in pulse 51945), RT2 ch. calibration no. 9 and 16 with -4.096 V | OK
519542025-03-17 14:13:40.659rhenriques000Repeat with changed ADC module on ch. 28 | OK, MARTe acquired this time
519532025-03-17 11:21:51.019rhenriques000Informatic pulse to test B3 (2MSps) after swapping chs. 22<->23 and 27<->28 and calibrate RT2 channels: 9 and 16 with 4.096 V | MARTe acquistion failed (operator fault)