Session aimTest MARTe and PF coils, then peform a series of plasma pulses and Spectrometer calibration pulse
Session day2025-01-29Operatorfernando
Session IDISTTOK_2025129_1Coordinatorrhenriques

Shot Time Operator Iplasma(kA) n(1018m-3) Duration (ms) Pre/post shot comments
516842025-01-29 17:10:59.101rhenriques000Calibrate spectrometer with Hg light bulb, C0=0, C1=1, C2=0, C3=0 | Good
516832025-01-29 16:30:33.651rhenriques000Trying again this time with DNS server set | Acquisition storage in SDAS successful. Signal name is POST.PROCESSED.TEST
516822025-01-29 15:40:59.898rhenriques000Trying again | Acquisition successful but it did not store data in SDAS, because the computer had no DNS resolver
516812025-01-29 15:36:57.534rhenriques000Run informatic pulse to test acquistion and triggering of SARSPEC spectrometer | Forgot to set the spectrometer ready for trigger
516802025-01-29 14:18:49.968rhenriques000Repeat with fixed cabling | OK, oscillations are seen in both sine and cosine signals
516792025-01-29 14:14:48.228rhenriques000Repeat with oscillations in the interferometer to induce fake phase shifts | OK, cosine amplitude is low, one cable at interferometer seemed slightly disconnected
516782025-01-29 13:48:11.862rhenriques000Informatic pulse to test MARTe and FireSignal communication | OK