Session aimTest ATCA acquisition modules
Session day2025-01-07Operatorfernando
Session IDISTTOK_202517_1Coordinatorrhenriques

Shot Time Operator Iplasma(kA) n(1018m-3) Duration (ms) Pre/post shot comments
516562025-01-07 18:28:11.594fernando-0.500Repeat with more gas | Better, plasma current spike at ~76 ms
516552025-01-07 18:23:18.583rhenriques000Attempting plasma to get light for the optical diganostics tests, added toroidal current signal to be acquired in MARTe RT2ch5 | No plasma, maybe too low pressure
516542025-01-07 17:58:55.663fernando000Repeat with Sine wave | Looks good again
516532025-01-07 17:54:49.939fernando000Repeat with square wave of 14.5 Vpp for calibration and check the signal distortion | Looks good
516522025-01-07 17:31:44.825fernando000Repeat with ~32 Vpp and lower frequency | clear saturation observed
516512025-01-07 17:19:17.789fernando000Repeat with signal cable connected to ATCA | No clear saturation observed
516502025-01-07 17:14:31.09fernando000Repeat with ~29 Vpp and smaller frequency to get a clear saturation | Forgot to connect cable to data acquistion...
516492025-01-07 17:06:08.65fernando000Test input range of acq. modules for MARTe: RT2 ch.17-24 with ~22 Vpp | Does not seem to saturate...
516482025-01-07 16:24:22.737fernando000Repeat with ~27 Vpp input tringular waveform signal | saturation observed, need to check exact saturation amplitude (likely to be 25 Vpp)
516472025-01-07 16:19:29.777rhenriques000Triangular wave with 22 Vpp in B3 ch9 to ch16 | No saturation observed, input range of acq. modules is > 22 Vpp